• No : 6049
  • Date of Issue : 2021/05/18 15:33
  • Last Updated : 2021/06/11 14:27
  • Print

AOCm can not start[Error Code:6004 6026]

Start is attempted under the following conditions:
• Door is open.
• Error is present.
• Power was turned on with the door open, and the [RESET] button is not pushed.
• The syringe replacement procedure has not been completed.
• The turret type and the GC sampler presence setting are not consistent.
Category : 

Answer

Close the door, remove the cause, and press the [RESET] button.
If the error occurs again after taking the measures, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.