Replace the insert and then reset the counter.
Insert replacement
Counter reset
<GC instrument monitor setting>
Press HOME button > [Function] > [Diagnosis] > [Analysis counter],
and then display [Analysis counter] screen and reset the liner usage counter.
<Labsolutions setting>
Press [System Check]> [Consumables Reset] on the assistant bar, and reset the counter from the [Reset] button.