(EDX) No spectrum name is identified. The intensity indication is not correct.
(EDX) No spectrum name is identified. The intensity indication is not correct.
Category :
Products Search
>
Elemental Analysis
>
Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
Answer
Check Points: Use the sample for energy calibration, and check if peaks appear at the correct positions.
Remedies: Perform energy calibration. Perform an energy check.