• No : 3015
  • Date of Issue : 2020/11/09 15:36
  • Print

How to set the exposure time and how to reset the total lamp usage time for the PDA detector

How to set the exposure time and how to reset the total lamp usage time for the PDA detector.
Category : 

Answer

1. If the LabSolutions analysis program is running, exit the analysis program.
2. Click the (Start) menu to display the list of applications and then click [LabSolutions]-[Instrument Maintenance Tools].
3. Select the desired options in the [Tool Launcher] window.
   Select [xxx Utility] for [Application]. (For xxx Utility, select the name with the name of the device you want to maintain.)
   Select the analysis instrument that is connected to the instrument for maintenance for [Detector Name].
4. Select a registered username from the [User ID] list, enter the password, and click [OK].
5. [xxx Utility] is displayed.
 
6. Click the (Wavelength Calibration) button in the [PDA Utility] sub-window.
7. Exposure Time
   a) Select [Set Exposure Time] in the [Function] menu.
   b) The exposure time is automatically set, and the result is displayed.
   c) Check the exposure time value, and click [Yes].
   d) The exposure time and the voltage of the tungsten (W) lamp are set.
8. Reset the Total Lamp Usage Time
   a) Select [Lamp Time] in the [Function] menu.
   b) The total usage time of each lamp is displayed.
   c) Clicking [Reset] for each lamp resets the total usage time.