Cleaning the Needle

Cleaning the needle with a rinse kit.
Category : 

Answer

When you receive an error message that the needle is clogged, it may be clogged with salt or high boiling point components from a salting sample or high temperature analysis. Cleaning with a rinse kit may remove blockage.
 
Needle Rinse Kit (221-86643-41)
PN(Parts Number) Name Qty
228-15672-91 Adaptor for syringe  1
046-00043-05 5mL Syringe  1
228-16004-01
Coupling
1
 
For instructions on use, please refer to the attached instruction manual.
Attached file : 
221-86642A_needle_rinse_kit.pdf