• Font Size
  • S
  • M
  • L
  • No : 8692
  • Date of Issue : 2022/05/23 13:35
  • Print

(SMX-225CT) Fluoroscopic images are blurry

(SMX-225CT) Fluoroscopic images are blurry
Category : 

Answer

Cause 1: Are the X-ray focus setting values appropriate?
Corrective Measures 1: Adjust the focus values.
  Click here for "(SMX-225CT) Replacing the Filament - focus".
 
Cause 2: Is the target rotated regularly?
Corrective Measures 2: Rotate the target.
  Click here for "(SMX-225CT) Rotating the X-ray Tube Head Tungsten Target".
 
Cause 3: Have high voltage cables been greased on a regular basis?
Corrective Measures 3: Contact your system manager to have the high voltage cables greased by a Shimadzu sales or service representative.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.