• Font Size
  • S
  • M
  • L
  • No : 8691
  • Date of Issue : 2022/05/23 13:30
  • Print

(SMX-225CT) The live image remains completely black

(SMX-225CT) The live image remains completely black
Category : 

Answer

Cause 1: Are X-rays being emitted?
Corrective Measures 1: Emit X-rays.
 
Cause 2: Is the [OPERATE] key switch on the operation box turned to [OFF]?
Corrective Measures 2: Turn the [OPERATE] key switch on the operation box to [ON] to restart the CT software inspeXio64.
  Click here for "(SMX-225CT) Supplying Power to the System".
 
Cause 3: Is the signal cable disconnected between Flat panel detector and control computer?
Corrective Measures 3: Make sure the signal cable between Flat panel detector and control computer is securely connected to the image capture board in the control computer, then restart the CT software inspeXio64.
 
Cause 4: Is the kV or μA monitor value in the X-ray control area in the CT software inspeXio64 too low?
Corrective Measures 4: Increase the voltage or current setting.
 
Cause 5: Are the live image quality settings in the CT software inspeXio64 set appropriately?
Corrective Measures 5: Click the [AW] (Auto Window) button in the live image quality settings area in the CT software inspeXio64.
  

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.