(SMX-225CT) System does not start even after turning the [OPERATE] key switch
(SMX-225CT) System does not start even after turning the [OPERATE] key switch
Answer
Cause 1: Is the power supply circuit breaker at the site power supply service panel or main power supply circuit breaker on the back of the unit switched OFF?
Corrective Measures 1: Switch the site or system power supply circuit breaker ON.
Click here for "(SMX-225CT) Supplying Power to the System"
Cause 2: Is the operation box cable connected with the operation box normally?
Corrective Measures 2: Connect an operation box cable with an operation box certainly.Connect an operation box cable with an operation box certainly.
Click here for "(SMX-225CT) Restoring the System After an Error Shutdown".
Cause 3: Is the [EMERGENCY STOP] button on the operation box locked in the down position?
Corrective Measures 3: Unlock the [EMERGENCY STOP] button on the operation box by turning the button clockwise.