FAQ - Frequently Asked Questions
- Products Search > Non-Destructive Testing (NDT) Systems > inspeXio SMX-225CT > (SMX-225CT) Error Detection on Flat Panel Detector
(SMX-225CT) Error Detection on Flat Panel Detector
(SMX-225CT) Error Detection on Flat Panel Detector
Answer
If any abnormality occurs on a device that captures image data, such as the flat panel detector and the command processor which retrieves image data from the flat panel detector, the following error messages will appear.
If this error message is displayed, shut down the system and CT software according to "Shutting Down the System", and restart them according to "Supplying Power to the System" and "Starting the Software". If the error message repeatedly appears after restarting, contact contact a Shimadzu sales or service representative to report the details of the message.
If this error message is displayed, shut down the system and CT software according to "Shutting Down the System", and restart them according to "Supplying Power to the System" and "Starting the Software". If the error message repeatedly appears after restarting, contact contact a Shimadzu sales or service representative to report the details of the message.
Click here for "(SMX-225CT) Shutting Down the System".
Click here for "(SMX-225CT) Supplying Power to the System".
Click here for "(SMX-225CT) Starting the Software".
01: Receptor Config Error.
10: Wrong Command Format Received by network.
12: Ethernet Control Error.
16: msgQSend Call Failed.
24: FLASH File Error.
25: Memory Access Error.
30: System control Error.
32: Receptor Error.
34: Receptor Medium Low.
35: Receptor Medium High.
37: Receptor Firmware Error.
43: Link Ready Error Detected.
45: ECP Clock is NOT set.
A0: FPGA temperature Warning.
A1: FPGA temperature Error.
A2: FAN1 Error.
A3: FAN2 Error.
A4: FAN1 and FAN2 Error.
A5: Too many deficits.
A6: Backup Boot.
A7: System control unavailable.
E4: Failed BOOT test.
10: Wrong Command Format Received by network.
12: Ethernet Control Error.
16: msgQSend Call Failed.
24: FLASH File Error.
25: Memory Access Error.
30: System control Error.
32: Receptor Error.
34: Receptor Medium Low.
35: Receptor Medium High.
37: Receptor Firmware Error.
43: Link Ready Error Detected.
45: ECP Clock is NOT set.
A0: FPGA temperature Warning.
A1: FPGA temperature Error.
A2: FAN1 Error.
A3: FAN2 Error.
A4: FAN1 and FAN2 Error.
A5: Too many deficits.
A6: Backup Boot.
A7: System control unavailable.
E4: Failed BOOT test.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.