(LC) Calibrating the pH Electrode

(LC) Calibrating the pH Electrode
 
Please select a maintenance item.
 
NOTE:
• When using a new pH electrode for the first time or if a pH electrode is not used for a long period of time (one week or longer), the dry pH responsive glass membrane may cause delays in response. Therefore, after rinsing the cell, fill the cell with purified water or pH7 standard solution with the refilling port open and leave it for half a day to a whole day.
• You must complete calibration before starting analysis.