(MCE) Maintenance of Piercing Needle

(MCE) Maintenance of Piercing Needle
Category : 

Answer

« Purpose »
  To remove any substances adhered to the tip of the piercing needle.
  Adhered substances may prevent the needle from piercing the aluminum seal on the sample plate, or cause contamination of the sample.
 
« Guidelines for Procedure »
  When adhered substances are present on the tip of the piercing needle.
 
« Requirements »
  • Tweezers
  • Wipes
 
« Cleaning Procedure »
1. Perform the procedure in "Moving All Axes to Home Position".
  Click here for "(MCE) Moving All Axes to Home Position".
2. Open the top cover.
3. Using the tweezers, peel off any adhered substances on the tip of the piercing needle in a downwards direction.
  
4. Use the wipes to remove any adhered substances from the tweezers.