• Font Size
  • S
  • M
  • L
  • No : 8089
  • Date of Issue : 2020/12/02 14:16
  • Print

(AGX-V) The test force display is "- - - -"

(AGX-V) The test force display is "- - - -"
Category : 

Answer

• The load cell is not recognized. Turn off the power, check that the calibration cable is connected appropriately, and turn on the power again.
• If the type of display is "peak value" or "break point value", the value is displayed only after a test ends.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.