(AGX-V) Maintaining the Guide Poles
(AGX-V) Maintaining the Guide Poles
- Category :
Answer
For "table-top" testers, regularly grease the guide poles.
• Grease name: Molybdenum disulfide lithium-based grease
• Grease name: Molybdenum disulfide lithium-based grease
Part Number: 017-27014-02 GREASE,MO-S2 LG NO.2
• Greasing frequency: Every 6 months
HINT: Greasing is not necessary for "floor type".
• Greasing frequency: Every 6 months
HINT: Greasing is not necessary for "floor type".
Regularly grease the guide bush (2) of the crosshead (1) to keep the point (3) lubricated.
NOTE: Turn off the power before opening the protection cover of ball screws (4).
Also, be sure to close the ball screw protection cover before turning on the power.
NOTE: Turn off the power before opening the protection cover of ball screws (4).
Also, be sure to close the ball screw protection cover before turning on the power.