(AGX-V) Maintaining the Ball Screws
(AGX-V) Maintaining the Ball Screws
- Category :
Answer
Regularly grease the ball screws to keep them lubricated.
• Grease name: General grease NLGI No. 2 (Shell Alvania Grease No. 2)
• Grease name: General grease NLGI No. 2 (Shell Alvania Grease No. 2)
Part Number: 339-89306-21 CARTRIDGE GREASE400CC
Part Number: 339-89108-01 GREASE CAN CH-400
• Greasing frequency: Every 6 monthsGrease the grease nipples (1) located on the ball screw nuts of the crosshead with a grease gun.
• Greasing frequency: Every 6 monthsGrease the grease nipples (1) located on the ball screw nuts of the crosshead with a grease gun.
Grease nipple positions