- Products Search > Materials Testing & Inspection > Autograph(AG) > AGX-V Series > (AGX-V) Checking the Stroke Speed
(AGX-V) Checking the Stroke Speed
(AGX-V) Checking the Stroke Speed
- Category :
Answer
1. Press the manual button on the smart controller to turn on the manual mode and record the current jog speed.
2. On the [Jog speed] setting screen, set the jog speed to 10mm/min.
Click here for "(AGX-V) Moving the Crosshead".
3. On the main screen, press (zero reset) of [Stroke] to set the stroke value to 0.
2. On the [Jog speed] setting screen, set the jog speed to 10mm/min.
Click here for "(AGX-V) Moving the Crosshead".
3. On the main screen, press (zero reset) of [Stroke] to set the stroke value to 0.
4. Start the stopwatch when the stroke value becomes 5mm and stop it when the stroke value becomes 15mm.
5. Check the time measured with the stopwatch. The normal value is 60±0.26 seconds.
HINT: The above value is based on the assumption that the error of stopwatch operation is 0.2 seconds.
6. On the [Jog speed] setting screen, return the jog speed to an original value.
5. Check the time measured with the stopwatch. The normal value is 60±0.26 seconds.
HINT: The above value is based on the assumption that the error of stopwatch operation is 0.2 seconds.
6. On the [Jog speed] setting screen, return the jog speed to an original value.
If the speed is abnormal, contact a Shimadzu sales or service representative.