If an alarm occurs in the AGS-X unit , an alarm sound i s emitted and the alarm screen flashes.
To cancel the alarm, eliminate the cause and press the [STOP] button.
NOTE: If a computer is connected, alarms cannot be canceled via the operating panel.
To cancel the alarm, click the [Reset] button displayed on the computer screen.
« Alarm Code »
Alarm Code 1: EMG Switch On
Description: The emergency stop button is pressed.
Corrective Measures: After confirming safety, release the emergency stop button.
To release the emergency stop button, turn the movable portion in the direction indicated by arrows or pull it up
Alarm Code 2: Cross Head Limit
Description: Crosshead limit was activated.
Corrective Measures: Temporarily loosen the crosshead limit collar and move the limit bar back to its freed position. Use jog operations to move the crosshead to a safe position.
Click here for "(AGS-X) Setting Crosshead Stroke Limits".
Alarm Code 10: Servo Off
Description: Crosshead movement was attempted with the servo OFF.
Corrective Measures: Confirm whether the emergency stop button or a crosshead limit is active.
Click here for "(AGS-X) Setting Crosshead Stroke Limits".
Alarm Code 11: Servo ALM
Description: An error has occurred in the servo amplifier.
Corrective Measures: The servo motor could be overloaded or the power supply voltage may have dropped. Switch the AGS-X power OFF and switch it ON again after about 10 seconds. If this does not resolve the problem, there may be a hardware failure. Contact
a Shimadzu sales or service representative.
Alarm Code 12: USB Func (PC) Err.
Description: An error occurred during USB communications with the computer.
Corrective Measures: Switch the AGS-X power OFF, then switch it ON again after about 10 seconds. Also close and restart the materials testing software on the computer.
Alarm Code 16: Inter Lock ALM
Description: The interlock for the fragment containment cover was activated.
Corrective Measures: If this alarm occurred when starting a test. The fragment containment cover is open. Close the fragment containment cover when starting a test. Also, the fragment containment cover must be opened and closed after finishing one test before the next test can be started.
If this alarm occurredd uring a test. The fragment containment cover was opened during the test. Do not open the containment cover during tests.
Alarm Code 18: PC Call Err.
Description: An error has occurred in the computer link function.
Corrective Measures: Switch the AGS-X power OFF, then switch it ON again after about 10 seconds. Also close and restart the materials testing software on the computer.
Alarm Code 20: Over Rating ALM
Description: Test force has exceeded +102% of rated load cell capacity.
Corrective Measures: Use load cells within their rated capacity. Jig weight should be within ±20% of the load cell rating and zero the test force with the jig attached.
Click here for "(AGS-X) Attaching a Tensile Test Jig".
Click here for "(AGS-X) Attaching Compression and Bending Test Jigs".
Alarm Code 21: Under Rating ALM
Description: Test force has exceeded 102% of rated load cell capacity.
Corrective Measures: Use load cells within their rated capacity. Jig weight should be within ±20% of the load cell rating and zero the test force with the jig attached.
Click here for "(AGS-X) Attaching a Tensile Test Jig".
Click here for "(AGS-X) Attaching Compression and Bending Test Jigs".
Alarm Code 22: Soft Limit ALM
Description: A software limit was triggered.
Corrective Measures: Confirm the software limit settings and whether they are enabled or disabled. If the settings are set properly, then use the system within the setting limits.
Alarm Code 23: Out of Control
Description: Stroke speed exceeded 500 mm/min during automatic contr ol mode, or the required automatic control parameter was determined incorrectly.
Corrective Measures: Confirm whether the specimen was attached properly to the jig. Also, if the initial speed is too high, a sudden increase in test force can cause the contr ol system to stop the test. Reduce the initial speed.
Click here for "(AGS-X) Attaching Specimens (Bending Test)".
Alarm Code 24: Touch Load ALM
Description: The contact load function was activated.
Corrective Measures: An obstruction was detected while the crosshead is being returned or is in jog operations. Do not touch the load cell or jig during return or jog operations.
Alarm Code 30: CalCable Err. (Read)
Description: Failed to read data from the calibration cable.
Corrective Measures: Confirm whether the calibration cable is connected correctly to the AGS-X unit . Also, switch the AGS-X power OFF and switch it ON again after about 10 seconds.
Alarm Code 31: CalCable Err. (Write)
Description: Failed to save data to the calibration cable.
Corrective Measures: Confirm whether the calibration cable is connected correctly to the AGS-X unit. Also, switch the AGS-X power OFF and switch it ON again after about 10 seconds.
Alarm Code 33: AutoZero Err.
Description: Auto zero of test force or sensor input failed.
Corrective Measures: Confirm whether the load cell cable and sensor input cables are connected correctly to the AGS-X unit. The range within which sensor data can be auto zeroed differs depending on the sensor type, but generally is within ±20% of full scale. If it exceeds this range, decrease the sensor input.
Alarm Code 34: F-CAL Err.
Description: Calibration of measured test force failed.
Corrective Measures: Confirm whether the load direction (tension/compression) is correct. Also, confirm whet her the correct load cell capacity is registered in the calibration cable. Confirm whether there was an error in the F-CAL procedure as well. If the error continues after repeated attempts, it could be a hardware failure. Contact
a Shimadzu sales or service representative.
Alarm Code 35: E-CAL Make Err.
Description: Creation of electronic test force calibration signal failed.
Corrective Measures: The load cell must be unloaded when the E-CAL signal is created. Eliminate any loads, then try again. Also, confirm whether the correct load cell capacity is registered in the calibration cable.
Click here for "(AGS-X) Calibrating Test Force (E-CAL)".
Alarm Code 36: E-CAL Err.
Description: Electronic calibration of test force failed.
Corrective Measures: The load cell must be unloaded to perform E-CAL. Eliminate any loads, then try again. Also, F-CAL must already be completed as well. If the error continues after repeated attempts, it could be a hardware failure. Contact
a Shimadzu sales or service representative.
Click here for "(AGS-X) Calibrating Test Force (E-CAL)".
Alarm Code 37: Linearizing Err.
Description: Creation of nonlinear correction factor failed.
Corrective Measures: Confirm whether the load direction (tension/compression) is correct. Also, confirm whether the correct load cell capacity is registered in the calibration cable. Confirm whether there was an error in the linearization procedure as well.
Alarm Code 38: Calibration Err.
Description: Calibration of analog input failed.
Corrective Measures: If the alarm occurred during zero adjustment. Confirm whether zero volts is applied to the analog input terminal.
If the alarm occurred during span adjustment. Confirm whether 5V or 10V is applied to the analog input terminal.
Alarm Code 39: E-CAL NonExe. ALM
Description: Attempted to start test without performing E-CAL after switching power ON or after connecting calibration cable.
Corrective Measures: Perform E-CAL.
Click here for "(AGS-X) Calibrating Test Force (E-CAL)".
Alarm Code 60: Parameter Err. (1)
Description: The method used for the single test or cycle test is incorrect.
Corrective Measures: Confirm the test speed and test direction.
Alarm Code 61: Parameter Err. (2)
Description: The test method for test force control, strain control, or stroke control is incorrect.
Corrective Measures: Confirm automatic control parameters.
Alarm Code 63: C.H. Return Err.
Description: Crosshead return was attempted without resetting position after changing test direction or test force polarity.
Corrective Measures: Reset the position whenever the test direction or test force polarity is changed.
Alarm Code 70: DataLety Err.
Description: Dataletty system does not enter ready mode.
Corrective Measures: Confirm whether the Dataletty unit is online.
Alarm Code 9999: System Err.
Description: An error other than those above occurred.
Corrective Measures: Switch the AGS-X power OFF, then switch it ON again after about 10 seconds. If the error continues after repeated attempts, it could be a hardware failure. Contact
a Shimadzu sales or service representative.