  • No : 7864
  • Date of Issue : 2022/01/24 14:12
  • Print

(LCMS) Checking the Status before or during Data Acquisition

(LCMS) Checking the Status before or during Data Acquisition
Category : 

Answer

The status of [Before acquisition] is for LCMS-2050 and 9030.
 
Data acquired with the MS unit can be used to check the temperature, gas flowrate, and other MS system status information before or during data acquisition.
Open the data file in the [MS Data Analysis] window.
On the [View] menu, click [Status of Data Acquisition].
 
Select the [Type] checkbox for the type of information to view. If [Before acquisition] is selected, the status can be checked up to 72 hours before acquisition.