(LCMS) Stopping a Realtime Batch

(LCMS) Stopping a Realtime Batch
Category : 

Answer

There are several ways to stop a realtime batch.
Click [Stop] on the [Realtime Batch] assistant bar during a realtime batch to display the following sub-window.
The operations change as follows depending on the combination of checked and cleared checkboxes.
  Data acquisition under execution: Selected、Batch Processing: Selected、Data acquisition stops for the row being executed, and is not executed for the subsequent rows. (Default operation)
  Data acquisition under execution: Selected、Batch Processing: Not selected、Data acquisition stops for the row being executed, and starts for the next row.
  Data acquisition under execution: Not selected、Batch Processing: Selected、Data acquisition is completed for the row being executed, but is not executed for the subsequent rows.
  Data acquisition under execution: Not selected、Batch Processing: Not selected、Invalid
 
HINT: When batch processing is stopped midway, the [>>] mark is displayed to the right of the row number of the next row in the batch table (the row where it would stop if a data file has not been made).
HINT: When batch processing is started, clicking the [Start Realtime Batch] icon on the [Realtime Batch] assistant bar displays a confirmation message asking whether to start on the next row in the batch table, or to start from the first row. Select one of these options.