FAQ - Frequently Asked Questions
- Products Search > Materials Testing & Inspection > Autograph(AG) > EZ-Test Series > (EZ-Test) Fuses fail frequently
(EZ-Test) Fuses fail frequently
(EZ-Test) Fuses fail frequently
- Category :
Answer
• Confirm whether the input voltage mode switch is set properly.
• Confirm whether the fuses current rating is correct. (100V: 10A, 220V: 5A)
• Confirm whether a time delayed type of fuse is being used. (A "fast acting" fuse cannot be used.)
• Confirm whether the fuses current rating is correct. (100V: 10A, 220V: 5A)
• Confirm whether a time delayed type of fuse is being used. (A "fast acting" fuse cannot be used.)
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.