(EZ-Test) Calibrating Test Force (E-CAL)

(EZ-Test) Calibrating Test Force (E-CAL)
Category : 

Answer

Electronically calibrates (E-CAL) the test force. Calibrating the test force with E-CAL uses precision resistors built into the calibration cable to apply a simulated load on the load cell and calibrate the test force amplifier
 
[[Setting Procedure]]
 
[Setting Value]
  Force: Electronically calibrates (E CAL) the test force.
 
CAUTION: Calibrate test force by E-CAL with the load cell in the unloaded state.
  Wait at least 15 minutes after switching the load cell power ON, before calibrating test force by E-CAL.
 
NOTE: E-CAL takes about 30 seconds. Please wait without operation.