• Font Size
  • S
  • M
  • L
  • No : 7523
  • Date of Issue : 2021/12/06 11:48
  • Print

(DUH) Indentation shape is abnormal

(DUH) Indentation shape is abnormal
DUH-211,211S,210,210S
Category : 

Answer

If the Indentation shape is abnormal as shown in below, the indenter mounting part may be defective. Replace the defective part with a new one. For replacement, contact a Shimadzu sales or service representative.
 
When a sample surface is not horizontally mounted, the indentation shape also becomes abnormal as shown in below. Make sure that the sample stage is leveled. Then, indent a reference sample, and check the indentation shape.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.