(DUH) Magnification Adjustment

(DUH) Magnification Adjustment
After removing the optical head or replacing the x50 objective lens, or when magnification becomes improper, perform the magnification adjustment procedure.
For the magnification adjustment, prepare the following equipment:
• Part Number: 046-60201-02  Objective micrometer (0.01 mm, Option)
• Hexagon wrench
 
When an optical head is exchanged, please perform rough adjustment and fine adjustment. The magnification adjustment by exchange of an object lens should fine adjustment.
 