FAQ - Frequently Asked Questions
- Products Search > Materials Testing & Inspection > Hardness Tester > Micro Hardness Tester(HMV) > (HMV) Troubleshooting (HMV-G-MM)
(HMV) Troubleshooting (HMV-G-MM)
(HMV) Troubleshooting (HMV-G-MM)
HMV-G-MM: Motorized Micrometer Kit
Answer
Problem1: The program does not start up.
Cause1: Part of the program was deleted, or the folder was changed.
Solution1: Reinstall the program.
Problem2: The initialization of the X-Y stage does not complete.
Cause2-1: The grease on the motorized micrometer is hardened due to long disuse.
Solution2-2: Soften the hardened grease and apply it evenly.
Click here for "(HMV) Maintenance and Inspection (HMV-G-MM)".
Cause2-2: Although the grease is not hardened, a step-out has occurred on the motor of the micrometer.
Solution2-2: Replace the motorized micrometer. Contact a Shimadzu sales or service representative.
Cause2-3: The motorized micrometer is not energized. (The knob is as tight as in the power-off status.)
Solution2-3: Replace the motor driver. Contact a Shimadzu sales or service representative.
Problem3: A stage connection error occurs.
Cause3: The cable is disconnected or almost disconnected.
Solution3: Confirm the cable connection.
Problem4: The stage does not move towards indicated directions.
Cause4-1: The stage has reached the limit of its movable range.
Solution4-1: Modify the sample position so that the testing position is located within the movable range.
Cause4-2: The power to the controller BOX was turned OFF.
Solution4-2: Turn ON the power.
Solution4-2: Turn ON the power.
Cause4-3: The cable is disconnected or almost disconnected.
Solution4-3: Confirm the cable connection.
Solution4-3: Confirm the cable connection.
Problem5: An alarm was displayed on the software window.
Cause5: The emergency stop button was pressed.
Solution5: Confirm that there is no abnormality on the instrument and its vicinity, and release the emergency stop button.
Click here for "(HMV) In an Emergency".
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.