(SALD) Cyclone Injection Type Dry Measurement Unit(SALD-DS5S), Troubleshooting

(SALD) Cyclone Injection Type Dry Measurement Unit(SALD-DS5S), Troubleshooting
SALD-DS5S: Cyclone Injection Type Dry Measurement Unit for SALD-2300
Category : 

Answer

Trouble1: Abnormality in measured light intensity data
Cause1: Sample is adhered to lens and/or filters
Counter measures1: Clean the measuring chamber
 
 
Trouble2: Air valve does not open
Cause2-1: Filter of dust collector is plugged
Counter measures2-1: Clean the dust collector filter
 
Cause2-2: Disconnection of cables and/or tubes
Counter measures2-2: Connect correctly
 
Cause2-3: Wrong insert position of the dust collector pipe
Counter measures2-3: Set the dust collector pipe inserting it to the upper one of the two receptacles
 
Cause2-4: Plugging of pressure switch filter and others
Counter measures2-4: Exchange the filter element and/or clean the silicone tube
  Click here for "(SALD) Dry measurement case(SALD-DC23), Replace the Filter element".
 
 
Trouble3: Frequent fuse cuttings
Cause3: Sampling unit Presser regulator unit
Counter measures3: Call the field engineer; parts may be deteriorated or in short circuit. Contact a Shimadzu sales or service representative.