• Font Size
  • S
  • M
  • L
  • No : 7453
  • Date of Issue : 2021/11/18 13:27
  • Print

(SALD) Sampler, Cleaning of the sample bath

(SALD) Sampler, Cleaning of the sample bath
MS-23 / MS-75
Category : 

Answer

WARNING: Never remove the main cover. Doing so may result in injury or malfunction.
Part Number: 347-62086  Sample bath seal
 
1. Release the lock nut of the feed pipe, and remove the feed pipe.
  
 
2. Release the screw of return pipe, and remove the return pipe.
  
 
3. Release the screw of the circulate pump axis.
  
 
4. Release the screws of the pump head.(2 screws)
  
 
5. Lift up the pump head and turn to the left side.
  
 
6. Release the screws of the sample bath.(3 screws)
  
 
7. Lift up and remove the sample bath.
  
 
8. Clean inside of the sample path. Use only water and a neutral detergent or ethanol.
 
9. After cleaning the sample bath, install the sample bath seal and sample bath.
  NOTE: Exchange the sample bath seal every time you disassemble it. Using old seal causes liquid leakage.
  
 
10. Assemble the remained parts.
 
11. Circulate liquid and check the leakage of liquid.
  Click here for "(SALD) Sampler, Check of the tubes(leak check)".

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.