(TOC) Setting the System Configuration (TOC-1000e)
(TOC) Setting the System Configuration (TOC-1000e)
Date/Time (clock setting), Screen brightness and button click sound, Standard text that you can select as the reason for change when changing settings (default comments), Day of the week and time to run a system check
- Category :
Answer
NOTE: When [User Authentication] on the [System Policy] screen is “Enabled,” the following operations are only available to an Administrator or a Main user level user. These operations are not available to User or Guest level users.
1. Press - [Settings].
2. Press [System].
• Setting the Date/Time
3. Press [Date & Time].
The settings screen appears.
Set the date and time (clock setting).
Set the date and time (clock setting).
(1) [Date]: Set this to the current date. Press the input field or to display a calendar.
(2) [Time]: Set this to the current time. Press the input field to display the number entry screen.
(3) [Date Format]: Set the date format. Select [YYYY / MM / DD], [MM / DD / YYYY], or [DD / MM / YYYY].
“YYYY” represents year, “MM” represents month, and “DD” represents date.
(4) [Apply]: Press this button to confirm the settings.
The set date/time is reflected in the date and time on the status bar displayed at the top of the screen.
(5) [Cancel]: Press this button to discard any changes made to the settings.
(3) [Date Format]: Set the date format. Select [YYYY / MM / DD], [MM / DD / YYYY], or [DD / MM / YYYY].
“YYYY” represents year, “MM” represents month, and “DD” represents date.
(4) [Apply]: Press this button to confirm the settings.
The set date/time is reflected in the date and time on the status bar displayed at the top of the screen.
(5) [Cancel]: Press this button to discard any changes made to the settings.
• Setting the Screen Brightness and Button Click Sound
3. Press [Screen Setting].
The settings screen appears.
Set the screen brightness and button click sound.
Set the screen brightness and button click sound.
(1) [Auto. Screen Off]: Set the time to wait before automatically turning off the screen display. Select [None] (No automatic turning off), [5 mins], [15 mins], or [30 mins].
To recover screen display after it has turned off, press any part on the screen.
(2) [Brightness]: Set the screen brightness.
(3) [Sounds] Set whether or not to make a click sound when a button on the screen is pressed.
Use the button to switch between Disable and Enable.
(4) [Apply]: Press this button to confirm the settings.
(5) [Cancel]: Press this button to discard any changes made to the settings.
To recover screen display after it has turned off, press any part on the screen.
(2) [Brightness]: Set the screen brightness.
(3) [Sounds] Set whether or not to make a click sound when a button on the screen is pressed.
Use the button to switch between Disable and Enable.
(4) [Apply]: Press this button to confirm the settings.
(5) [Cancel]: Press this button to discard any changes made to the settings.
• Setting the Default Comments
You can set standard text options (default comments) when the [Save Comment] setting on the [System Policy] screen is set to “Save Comment.”
3. Press [Default Comment Edit].
The settings screen appears.
Set the standard text that you can select as the reason for change when changing settings (default comments).
Set the standard text that you can select as the reason for change when changing settings (default comments).
(1) Comment list: Comments that are defined are displayed in this list.
NOTE: Up to 16 comments can be registered. The content can be changed, but no comment can be added or deleted.
(2) Cursor movement buttons: Use these buttons to move the selection cursor. The topmost display indicates the “Selected comment number / Total number of comments.”
(3) Edit button: Allows you to edit the selected comment.
(4) [Apply]: Press this button to confirm the settings.
(5) [Cancel]: Press this button to discard any changes made to the settings.
NOTE: Up to 16 comments can be registered. The content can be changed, but no comment can be added or deleted.
(2) Cursor movement buttons: Use these buttons to move the selection cursor. The topmost display indicates the “Selected comment number / Total number of comments.”
(3) Edit button: Allows you to edit the selected comment.
(4) [Apply]: Press this button to confirm the settings.
(5) [Cancel]: Press this button to discard any changes made to the settings.
• Setting the System Check
3. Press [System Check].
The settings screen appears.
Set the day of the week and time to run a system check.
Set the day of the week and time to run a system check.
(1) [Week of Month]: Set the week, the day of the week and time to run a system check.
(2) [Day of Week]: Set the week to run a system check.
(3) [Time]: Set the time to run a system check. Press the input field to display the number entry screen.
(4) [Apply]: Press this button to confirm the settings.
(5) [Cancel]: Press this button to discard any changes made to the settings.
(2) [Day of Week]: Set the week to run a system check.
(3) [Time]: Set the time to run a system check. Press the input field to display the number entry screen.
(4) [Apply]: Press this button to confirm the settings.
(5) [Cancel]: Press this button to discard any changes made to the settings.