(TOC) Actions to Take in Case of a Problem(Instrument alarm) (TOC-1000e)
(TOC) Actions to Take in Case of a Problem(Instrument alarm) (TOC-1000e)
- Category :
Answer
The operating status changes to [Warning], and
is displayed.
Press [OK] to return to the original screen.
• SAMPLE WATER: Sample liquid level is low.
→Make sure that the sample liquid is flowing into the TOC-1000e from the online port. *1
• BATTERY: Battery needs replacement.
→Contact your Shimadzu representative.
• BOARD TEMP.: The temperature inside the chamber of the instrument is outside of the rated operating range.
→Adjust the temperature around the instrument to be within the rated operating range. *1
→Adjust the temperature around the instrument to be within the rated operating range. *1
• SAMPLE TEMP.: The temperature of the sample liquid is outside of the rated operating range.
→Adjust the temperature of the sample liquid to be within the rated operating range. *1
→Adjust the temperature of the sample liquid to be within the rated operating range. *1
• PRINTER(OFFLINE): Printer is offline.
→Check whether the printer is turned on. Check the printer connection. *1*2
→Check whether the printer is turned on. Check the printer connection. *1*2
• PRINTER(PAPER OUT): The printer is out of paper.
→Check whether the printer has paper. *1*2
• EXCIMER LAMP(EXCHANGE): Reactor needs replacement.
→Turn off the power to the instrument and replace the reactor.
Click here for "Consumable Parts Replacement (TOC-1000e)".
• PUMP(EXCHANGE): Pump head needs replacement.
→Turn off the power to the instrument and replace the pump head.
→Turn off the power to the instrument and replace the pump head.
Click here for "Consumable Parts Replacement (TOC-1000e)".
• STORAGE OVER 80%: The instrument’s storage capacity usage has reached 80 %.
→Contact your Shimadzu representative.
→Contact your Shimadzu representative.
• REACTOR COVER: Lamp cover of the reactor is open.
→Open the front panel of the instrument and check whether the lamp cover of the reactor is closed. *1
→Open the front panel of the instrument and check whether the lamp cover of the reactor is closed. *1
*1: Alarms are automatically cleared once the cause is removed.
*2: Instrument alarms that are triggered by a problem with the printer are not printed from the printer even when the runtime report function is enabled.
*2: Instrument alarms that are triggered by a problem with the printer are not printed from the printer even when the runtime report function is enabled.
[Checking Whether an Instrument Alarm Has Been Cleared]
1. Press
at the top left of the screen.

2. Select the [Instrument Alarm] tab.

Instrument alarms cannot be cleared manually. Perform the above actions to remove the cause.
Some instrument alarms are automatically cleared once the cause is removed.
Some instrument alarms are automatically cleared once the cause is removed.