(TOC) Program Update (TOC-1000e)
(TOC) Program Update (TOC-1000e)
- Category :
Answer
NOTE: When [User Authentication] on the [System Policy] screen is “Enabled,” the following operations are only available to an Administrator level user. They are not operable by users of any other levels.
When a program update becomes necessary, update the program by performing the following steps.
1. Copy the update file to a USB flash drive.
Please get the update file from your Shimadzu representative.
Copy the file so that the "FwUpdate.smz" file is in the USB drive's root directory.
Please get the update file from your Shimadzu representative.
Copy the file so that the "FwUpdate.smz" file is in the USB drive's root directory.
2. Press - [Administration].
3. Press [Program Update].
The settings screen appears.
4. Remove the waterproof cover and connect a USB flash drive to the instrument.
5. Press [Check].
You will see the program version.
6. Press [Start].
The update process starts.
When it is completed, the instrument restarts on its own.
NOTE: Do not remove the USB flash drive until the restart is completed.
When it is completed, the instrument restarts on its own.
NOTE: Do not remove the USB flash drive until the restart is completed.