(TOC) Starting the Instrument (TOC-1000e)

(TOC) Starting the Instrument (TOC-1000e)
Category : 

Answer

NOTE: Before starting the instrument, make sure that the TOC-1000e and the optional units are correctly installed with correct cabling and piping.
 
1. Loosen the retaining screw and remove the waterproof cover.
  
2. Turn on the main switch on the left side.
  
  The Start button on the front side blinks, and the screen displays [Press the button].
3. Attach the waterproof cover.
4. Press the Start button on the front side.
  
  The initialization starts, and the startup screen is displayed.
  
  When the initialization is completed, the Home screen appears.
  
  You can start measurement from the Home screen.