• Font Size
  • S
  • M
  • L
  • No : 6823
  • Date of Issue : 2021/08/20 11:25
  • Print

(GAS) Gas does not flow into the flow meter (CGT-7100)

(GAS) Gas does not flow into the flow meter (CGT-7100)
Category : 

Answer

• There is a problem with the pump. → Replace the pump diaphragm or replace the pump completely. Contact a Shimadzu sales or service representative.
• There is a flow lag in the flow meter. → Disassemble and clean the flow meter. Contact a Shimadzu sales or service representative.
• Gas is leaking inside the analyzer. → Check the leak in the piping. Click here for "Gas leak checking (CGT-7100)".
• Drain has entered the mist absorber. → Replace the mist absorbent. Click here for "Maintenance (Replacement of consumables)".

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.