(GAS) The indicator values fluctuate or exhibit noise (CGT-7100)

(GAS) The indicator values fluctuate or exhibit noise (CGT-7100)
Category : 

Answer

• The ground is not complete, → Connect a ground having a resistance of no more than 100 ohms.
• There is a large source of vibration in the vicinity. → Remove the system from the vibration source or use anti vibration measures.
• The measurement cell is dirty. → Clean the cell. Contact a Shimadzu sales or service representative.