• Font Size
  • S
  • M
  • L
  • No : 6817
  • Date of Issue : 2021/08/20 11:25
  • Print

(GAS) Ozone odor is evident (NOA-7100)

(GAS) Ozone odor is evident (NOA-7100)
Category : 

Answer

Has temperature of ozone deactivator risen. (The ozone deactivator is normal if it is warm to the touch.)
 
NOTE: It is dangerous if the ozone deactivator is not replaced for a year or longer because highly concentrated ozone gas is exhausted from the exhaust ou tlet of the instrument. Replace the ozone deactivator once a year. And set the exhaust discharge area outdoor without applying any pressure to the exhaust outlet.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.