(ICPE) Response in Emergency
In the event of an emergency situation, such as a fire or an earthquake, if there is not enough time to stop the ICPE in the usual way, turn OFF the ICPE's power switch. The power supplied to the ICPE is interrupted.
Stop the supply of argon gas to the ICPE by closing the main valve of the argon gas supply.
Stop the supply of argon gas to the ICPE by closing the main valve of the argon gas supply.
Stop the supply of cooling water to the ICPE.
NOTE: If there is enough time, stop the ICPE as described "Stopping the ICPE in Daily Use"
NOTE: If there is enough time, stop the ICPE as described "Stopping the ICPE in Daily Use"
Click here for "Stopping the ICPE in Daily Use"
Answer
1. Recovery after Emergency Stop
After stopping the ICPE by turning OFF the power switch, perform the following recovery procedure.
1) Turn OFF the computer power switch.
2) Turn ON the power switch.
3) Start the ICPE as described in "Starting the ICPE".
2) Turn ON the power switch.
3) Start the ICPE as described in "Starting the ICPE".
2. Response to Power Failure
In the event of a power failure, perform the following procedure.
1) Turn OFF the ICPE's power switch and the computer power switch.
2) When power is restored, start the ICPE-9800 as described in "Starting the ICPE".
2) When power is restored, start the ICPE-9800 as described in "Starting the ICPE".
Click here for "Starting the ICPE in Daily Use".