(ICPE) Checking Operation of Safety Devices

(ICPE) Checking Operation of Safety Devices
Category : 

Answer

The operation of the safety devices used with the ICPE-9800 can be checked as described below. Check the operation of the safety devices on a regular basis (every one month).
 
1. Argon gas pressure sensor:
  1)Ignite the plasma.
  2)Stop the supply of argon gas and check that the plasma is extinguished.
 
2. Plasma stand door switch:
  1) Ignite the plasma.
  2) Open the plasma stand door and check that the plasma is extinguished.
 
3. Cooling water flow switch for RF coil, RF power supply, and orifice assembly:
  1) Ignite the plasma.
  2) Stop the supply of cooling water and check that the plasma is extinguished