• Font Size
  • S
  • M
  • L
  • No : 6532
  • Date of Issue : 2021/07/06 10:44
  • Last Updated : 2021/08/02 16:24
  • Print

(PDA) The spectrometer inner pressure has problem

PDA-8000: The spectrometer inner pressure is not within the green range.
PDA-7000: The Vacuum does not become OK.
Category : 

Answer

« Cause »
  • The O-ring of the condenser lens is not correctly installed.
  • The oil level in the vacuum pump is not the specified one.
  • The pump oil in the vacuum pump is soiled, lowering the exhaust ability.
 
« Measure »
  Perform maintenance and inspection of the condenser lens and vacuum pump.
  • Click here for Replacing the Pump Oil
 

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.