- Products Search > GC > GC-2030 > Error/Warning > AOC-20s up/down error[Error Code:6016]
AOC-20s up/down error[Error Code:6016]
The following causes are possible.
• The lifting motor has a problem.
• The deviation of the installation position from the teaching result is causing the gripper to hit the vial.
• There is an obstacle in the path to the destination, which is interfering with the vial.
• A vial with dimensions different from those of the vial used for teaching is used.
• The lifting motor has a problem.
• The deviation of the installation position from the teaching result is causing the gripper to hit the vial.
• There is an obstacle in the path to the destination, which is interfering with the vial.
• A vial with dimensions different from those of the vial used for teaching is used.
- Category :
Answer
Check and perform the following.
• Check whether the initialization process at power-on can be completed.
• Redo the teaching using the vial to be used for actual analysis.
• Check whether the initialization process at power-on can be completed.
• Redo the teaching using the vial to be used for actual analysis.
[In case of AOC-30i]
AOC-30i_AOC-20sU_SamplerTeaching.pdf
AOC-30i_AOC-20sU_SamplerTeaching.pdf
[In case of AOC-20i]
AOC-20i_AOC-20s_SamplerTeaching.pdf
• Perform a transport test (in case of AOC-30i).
• Perform a transport test (in case of AOC-30i).