AOC-20s rotating error[Error Code:6014]

The following causes are possible.
• The rotary motor has a problem.
• A vial with dimensions different from those of the vial used for teaching is used.
Category : 

Answer

Redo the teaching using thea vial to be used for actual analysis.
   [In case of AOC-30i]
      AOC-30i_AOC-20sU_SamplerTeaching.pdf
   [In case of AOC-20i]
      AOC-20i_AOC-20s_SamplerTeaching.pdf