AOCm sample vial is not set[Error Code:6008 6030]

The following causes are possible.
• No vial is placed in the position of the specified vial No.
• Vial is missing a vial cap.
• A vial with dimensions incompatible with the AOC is used.
• Vial is placed in a position exposed to direct sunlight.
• The vial sensor has a problem.
Category : 

Answer

[In case of AOC-30i]
• Check whether vials conforming to the equipment are being used.
 
 
• Check whether the operating environment is appropriate.
 
If the error appears even after checking the above,turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
 
[In case of AOC-20i]
Confirm that the vial to be analyzed is set. Also, make sure that the vial cap is on.
If the error appears even after checking the above,turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.