• Font Size
  • S
  • M
  • L
  • No : 5964
  • Date of Issue : 2021/05/18 15:37
  • Last Updated : 2021/06/04 10:26
  • Print

DET#n set temperature is low[Error Code:1049 1050 1051 1052]

Detector temperature is set at a lower value than column temperature.
Category : 

Answer

・To prevent detector contamination, set a detector temperature higher than the column oven temperature.
・When setting a detector temperature lower than the column oven temperature, turn off the contamination prevention function.
 
   ・[HOME] - [Function] - [Configuration] - [Max Temperature] to display [Max Temperature] screen.
   ・Display the second page and turn off the contamination prevention function of the relevant detector. 
 

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.