(EDX) Mylar film and "Mg" analysis

I have analyzed Mg(magnesium) in vegetation, but sufficient X-ray intensity was not obtained. For this analysis, a powder sample was placed in a sample container lined with a Mylar film. Is there any way to improve the intensity?
Category : 

Answer

If a Mylar film is used, the Mg intensity will be significantly reduced.
Either use a polypropylene film or a pressure molded briquette.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.