(EDX) Quantitative analysis with AsKα

In a qualitative analysis, As(arsenic) was detected.
I would like to quantify it using the AsKα line rather than using the AsKβ line.Is this possible?
Category : 

Answer

Yes, this is possible.
The analysis lines (Kα, Lα, etc.)used in quantitative calculations are set in the sensitivity coefficients method files (sysmetal, sysoxide)*. Change the analysis line in the sensitivity coefficients method file, and then recalculate the values.
* Sysmetal: Compound form "metal"
* Sysoxide: Compound form "oxide" 