(EDX) Quantitative calculation of FP method
With the FP method, how are quantitative values calculated?
Answer
With the X-ray fluorescence spectrometry, if the composition (elements and content) of the sample is known, X-ray fluorescence intensities can be calculated theoretically using measurement conditions and fundamental parameters (physical constants), based on the basic principles of fluorescent X-ray generation.
With the FP method, this theoretical intensity calculationis used to obtain the compositionfrom the measured intensity.
With the FP method, this theoretical intensity calculationis used to obtain the compositionfrom the measured intensity.