FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis > Energy Dispersive X-ray Fluorescence Spectroscopy(EDX) > (EDX) No spectrum name is identified. The intensity indic...
(EDX) No spectrum name is identified. The intensity indication is not correct.
(EDX) No spectrum name is identified. The intensity indication is not correct.
Answer
Check Points: Use the sample for energy calibration, and check if peaks appear at the correct positions.
Remedies: Perform energy calibration. Perform an energy check.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.