(EDX) No spectrum name is identified. The intensity indication is not correct.
(EDX) No spectrum name is identified. The intensity indication is not correct.
Answer
Check Points: Use the sample for energy calibration, and check if peaks appear at the correct positions.
Remedies: Perform energy calibration. Perform an energy check.