(EDX) X-Ray Tube Ageing
CAUTION: If the product has not been used for one month or longer, be sure to perform ageing of the X-ray tube.
Without X-ray tube ageing the service life of the X-ray tube could be shortened.
Without X-ray tube ageing the service life of the X-ray tube could be shortened.
If the instrument has not been used for one month or longer, X-ray tube ageing will be executed automatically when the software launches.
To ensure the safe and stable operation of the X-ray tube for a long time, ageing of the X-ray tube is necessary.
Only when ageing is finished, analysis can be started.
To ensure the safe and stable operation of the X-ray tube for a long time, ageing of the X-ray tube is necessary.
Only when ageing is finished, analysis can be started.