• Font Size
  • S
  • M
  • L
  • No : 4888
  • Displayed Date : 2021/01/25 11:36
  • Updated DTM : 2021/03/16 16:07
  • Print

(RF) Sipper6000 Rinsing Flow Cells

(RF) Sipper6000 Rinsing Flow Cells
Category : 

Answer

After measurement is completed, rinse the flow cell with distilled water or detergent each time. After rinsing the flow cell, turn OFF the power to the RF-6000 with the flow cell inside filled with distilled water. If the inside of the flow cell is dried, the inner wall of the cell may be contaminated with dirt, and as a result bubbles may easily occur.
To rinse the flow cell, start up the LabSolutions RF application and click the suction button on the measurement tool bar. The peristaltic pump will rotate to sip distilled water or detergent according to the set parameters.
 
For detergent, use liquid types as shown in the instruction manual.
NOTE: When using clinical samples, never forget to rinse the flow cell after daily use. Otherwise, fibrous impurities, often contained in such samples, may be solidified in the flow cell, and cause poor measurement repeatability.
 
Dilute concentration: Used in about 2 %

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "OKBIZ. for FAQ", which is a product of PKSHA Communication Inc.