• Font Size
  • S
  • M
  • L
  • No : 4750
  • Displayed Date : 2021/01/12 13:50
  • Updated DTM : 2021/01/12 13:51
  • Print

(HPLC) RID "OVER" is displayed

(HPLC) RID "OVER" is displayed
Category : 

Answer

• "OVER" is displayed in the "range" field of the initial screen.
  Cause: The recorder output has exceeded the scale limit.
  Actions: Press [zero].
 
• "OVER" is displayed in the "balance" field of the initial screen.
  Cause: Optical balance is shifted.
  Actions: Press [bal] to balance the optical system.
    * "OVER" may be displayed when peak of chromatogram is detected.
      This is not a problem. In this case, you need not balance the optical system.
    * Air bubbles may have entered in flow line and the optical balance may be shifted. Purge the flow line and balance the optical system.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "OKBIZ. for FAQ", which is a product of PRAZNA Inc.