• Font Size
  • S
  • M
  • L
  • No : 4537
  • Date of Issue : 2020/12/11 16:25
  • Print

(HPLC) Replace of Mixer 0.5-2.6mL, 100uL line filter

(HPLC) Replace of Mixer 0.5-2.6mL, 100uL line filter
Category : 

Answer

If the back pressure is more than 1MPa when the flowrate is 1mL/min and the mixer outlet is clogged. Remove the male nut and loosen the screw to replace the filter.
CAUTION:
a) Do not absolutely open the inner part.
b) Note that the specified performance may not be obtained if this unit is installed in a place subjected to vibration or fixed obliquely.
 
Parts Number: 228-18872-93 Filter
High Pressure Type: 228-18872-94 Filter,HP
The high pressure type is labeled with [P] as shown in the figure below.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.