Error Code:[0x5AD1]

Error Message: PDA detector: Initialization was not able to finish normally.
Category : 

Answer

«Main Cause»
Detector initialization failed.
 
«Countermeasure»
Implement corrective measures in the following sequence.
(1) Reconfirm that the flow cell is installed correctly.
(2) Since air bubbles may be contained in the flow cell, the mobile phase, water, methanol, etc. are passed and removed.
If that does not resolve the error, switch the instrument power OFF and contact a Shimadzu sales or service representative.