Error Code:[0x5AC4]

Error Message: PDA detector: W lamp does not light. Replace the W lamp.
Category : 

Answer

«Main Cause»
The W lamp failed to illuminate due to a broken filament or the illumination circuit is not functioning properly.
 
«Countermeasure»
Replace the W lamp.
Click here for Replacement method of Lamp (HPLC)
If replacing the W lamp does not resolve the error, switch the instrument power OFF and contact a Shimadzu sales or service representative.